CleanControlling Medical GmbH & Co. KGTests as part of the validation of instructions for the sterilization of medical devices (according
In the same way as with processing of medical devices (cleaning and disinfection) the manufacturer or the distributor must describe and provide a suitable sterilization method if the medical device is designated for sterilization.
To achieve the required sterility safety level the medical device manufacturer must indicate at least one validated sterilization method which meets the requirements of an international standard wherever possible, e.g. damp heat (ISO 17665).
To verify the suitability of the sterilization method indicated in the product-related instructions the sterile product is inoculated with the spores of very heat-resistant bacteria Geobacillus stearothermophilus. The inoculated product is packaged in accordance with the specifications (primary and, if applicable, secondary packaging) and then sterilized with steam with the specified parameters.
The steam sterilization is performed in a half-cycle process (e.g. 134°C, 2.5 minutes). The success of sterilization is then tested by inserting the product in a liquid culture medium for 7 days. The culture medium must not show any signs of bacteria growth, i.e. it must not be clouded.
Manufacturers and distributors of medical deviceswhich are designed for reuse, must provide information for their products and instructions for processing of the products in order to guarantee th...
In the same way as with processing of medical devices (cleaning and disinfection) the manufacturer or the distributor must describe and provide a suitable sterilization method if the medical device is...
Particulate contamination of the products is also becoming increasingly important in medical technology. Some questions however remain without sufficient answers. Exactly what damage do particles caus...
The ability of a substance to damage cells or tissue is referred to as cytotoxicity.
The test for cytotoxicity is one of the most important test methods within the scope of the biolog...
The objective of determination of bioburden is to determine the population of viable microorganisms (bacteria and fungi) on or in a product or its packaging.
The determination of this number is...
Endotoxins (ancient Greek endo: "inside", "within", toxins, "the poisonous substance") are part of the outer cell membrane of gram-negative bacteria, which can lead...
Hygiene is an important commodity at all times, but especially in the current crisis situation. Despite the fact that restrictions are beginning to be released, strict compliance with hygiene measures...